Title | Optomechanical reference accelerometer |
Publication Type | Journal Article |
Year of Publication | 2015 |
Authors | Gerberding, O, Cervantes, FGuzman, Melcher, J, Pratt, JR, Taylor, JM |
Journal | Metrologia |
Volume | 52 |
Issue | 5 |
Pages | 654 |
Date Published | 2015/09/08 |
Abstract | We present an optomechanical accelerometer with high dynamic range, high bandwidth and read-out noise levels below 8 ${\mu}$g/$\sqrt{\mathrm{Hz}}$. The straightforward assembly and low cost of our device make it a prime candidate for on-site reference calibrations and autonomous navigation. We present experimental data taken with a vacuum sealed, portable prototype and deduce the achieved bias stability and scale factor accuracy. Additionally, we present a comprehensive model of the device physics that we use to analyze the fundamental noise sources and accuracy limitations of such devices. |
URL | http://iopscience.iop.org/article/10.1088/0026-1394/52/5/654/meta;jsessionid=C2B417A5CD50B9B57EE14C78E1783802.ip-10-40-1-105 |
DOI | 10.1088/0026-1394/52/5/654 |